Elog Entry 2021-06-03: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 9: | Line 9: | ||
*Cannot reset HV for any of the detectors | *Cannot reset HV for any of the detectors | ||
*Walked Eric through power cycling MPS crate after the meeting | |||
*Was able to restore voltages for all PMTs and modules and enable all detectors | |||
[[File:ELBTPMT1B_PSfail.png|500px|PS fail]] | [[File:ELBTPMT1B_PSfail.png|500px|PS fail]] |
Latest revision as of 20:34, 3 June 2021
- Doing Post-Mortem tests after final changes have been implemented by Ricky
- This is after address change of modules was performed
- Starting 9:07 am
- Using CH2 on Module 1 first (ELBT:PMT1B)
- Enabling stop mode
- Tripped detector via LED at 250pC
- Then Hit diagnostics button and then saw power supply failed
- All HVs for all three modules tripped
- Cannot reset HV for any of the detectors
- Walked Eric through power cycling MPS crate after the meeting
- Was able to restore voltages for all PMTs and modules and enable all detectors